ClearPETTM

A high Performance Small Animal PET System


 
ClearPET  

Applications

 
 
PET   small animal

mouse, rat

rabbit, minipig

primate
 

 

 

Features

 
nuclide NaF bone marker
activity 300 µCi


Click on the picture to watch the movie (1.3MB)
  partial ring configuration

extendable gantry diameter

resolution 1.1 mm

sensitivity 4 %

multi modality with CT, MRI etc.
 
    Rat heart, 750 µCi FDG
   Rat heart, 750 µCi FDG
 
Detailed information

ClearPET brochure
The ClearPET PERFORMANCES
The ClearPET IMAGES
 
Gantry: crystal: LYSO & LuYAP in double layer
crystal size: (2 x 2 x 10) + (2 x 2 x 10) mm3
# crystal: 10240
# pixel: 5120
detector: 80 x Hamamatsu R7600M64
detector-Ø: 135 - 225 mm
animal port: 125 - 200 mm
axial length: 110 mm
 
Performance: timing resolution: 5.7ns (FWHM)
coincidence time: adjustable per software
field of view-Ø: adjustable 0 - 160 mm
energy resolution: < 30% LYSO & LuYAP
sensitivity: ∼ 5%
spatial resolution: 1.1 mm at center
≤ 2 mm, 30 mm off from center
modular system: 12, 16, 20 cassettes of gantry